Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-148
DMM7510-901-01 Rev. B / May 2015
:SYSTem:POSetup
This command selects the defaults that are used when you power on the instrument.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Nonvolatile memory
RST
Usage
:SYSTem:POSetup <name>
:SYSTem:POSetup?
<name>
Which setup to restore when you power on the instrument:
•
Power on to
*RST
defaults:
RST
•
Stored setup 0:
SAV0
•
Stored setup 1:
SAV1
•
Stored setup 2:
SAV2
•
Stored setup 3:
SAV3
•
Stored setup 4:
SAV4
Details
When you select
RST
, the instrument restores settings to their default values when the instrument is
powered on.
When you select a
SAV
option, the settings in the selected saved setup are applied when the
instrument is powered on. The settings are saved using the
*SAV
command.
Example
SYST:POS SAV1
Set the instrument to restore the settings
that are saved in the stored setup 1 when
the instrument is powered on.
Also see
(on page 6-14)