Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-138
DMM7510-901-01 Rev. B / May 2015
dmm.measure.bias.level
This attribute selects the amount of current the instrument sources when it makes measurements.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
1 mA
Usage
value
= dmm.measure.bias.level
dmm.measure.bias.level =
value
value
10
µ
A, 100
µ
A, 1 mA
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
Selects the amount of current that is sourced by the instrument to make measurements.
Example
dmm.measure.bias.level = 0.0001
Set a bias level of 100
µ
A.
Also see
dmm.measure.configlist.catalog()
This function returns the name of one measure configuration list that is stored on the instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
dmm.measure.configlist.catalog()
Details
You can use this command to retrieve the names of measure configuration lists that are stored in the
instrument.
This command returns one name each time you send it. This command returns
nil
to indicate that
there are no more names to return. If the command returns
nil
the first time you send it, no measure
configuration lists have been created for the instrument.