Section 3: Functions and features
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
3-90
DMM7510-901-01 Rev. B / May 2015
Trigger block summary
Front-panel
icon
SCPI command
TSP command
Block description
Not applicable
(on page 6-198)
(on page 8-284)
This returns the
settings for all
trigger model
blocks
(on page 6-181)
trigger.model.setblock() — trigger.BLOCK_BRANCH_ALWAYS
This defines a
trigger model block
that always goes to
a specific block
(on page 6-182)
trigger.model.setblock() — trigger.BLOCK_BRANCH_COUNTER
This defines a
trigger model block
that branches to a
specified block a
specified number of
times
Not applicable
:TRIGger:BLOCk:BRANch:COUNter:COUNt?
(on page 6-183)
trigger.model.getbranchcount()
(on page 8-285)
This returns the
count value of the
trigger model
counter block
:TRIGger:BLOCk:BRANch:COUNter:RESet
(on page 6-183)
trigger.model.setblock() — trigger.BLOCK_RESET_BRANCH_COUNT
(on page 8-324)
This creates a
block in the trigger
model that resets a
branch counter to 0
(on page 6-184)
trigger.model.setblock() — trigger.BLOCK_BRANCH_DELTA
This defines a
trigger model block
that goes to a
specified block if
the difference of
two measurements
meets preset
criteria
trigger.model.setblock() — trigger.BLOCK_BRANCH_ON_EVENT
(on
This branches to a
specified block
when a specified
trigger event occurs
:TRIGger:BLOCk:BRANch:LIMit:CONStant
(on page 6-187)
trigger.model.setblock() —
trigger.BLOCK_BRANCH_LIMIT_CONSTANT
This defines a
trigger model block
that goes to a
specified block if a
measurement
meets preset
criteria
:TRIGger:BLOCk:BRANch:LIMit:DYNamic
(on page 6-188)
trigger.model.setblock() —
trigger.BLOCK_BRANCH_LIMIT_DYNAMIC
This defines a
trigger model block
that goes to a
specified block in
the trigger model if
a measurement
meets user-defined
criteria