Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-136
DMM7510-901-01 Rev. B / May 2015
:SYSTem:BEEPer[:IMMediate]
This command generates an audible tone.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:BEEPer[:IMMediate] <frequency>, <duration>
<frequency>
The frequency of the beep (20 to 8000 Hz)
<duration>
The amount of time to play the tone (0.001 to 100 s)
Details
You can use the beeper of the instrument to provide an audible signal at a specific frequency and
time duration.
Using this function from a remote interface does not affect audible errors or key click settings that
were made from the Model DMM7510 front panel.
Example
:SYSTem:BEEPer 500, 1
Beep at 500 Hz for 1 s.
Also see
None
:SYSTem:CLEar
This command clears the event log.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:CLEar
Details
This command removes all events from the event log, including entries in the front-panel event log.
Also see