Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-119
Each index contains the settings for the selected function. Settings for other functions are not affected
when the configuration list index is recalled.
This command returns data for one configuration index.
Example
:SENSe:CONF:LIST:RECall "MyMeasList", 5
Recalls configuration index 5 in a configuration
list named
MyMeasList
.
:SENSe:CONF:LIST:RECall "MyMeasList"
Since an index was not specified, this command
recalls configuration index 1 from a configuration
list named
MyMeasList
.
Also see
(on page 3-37)
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 6-116)
[:SENSe[1]]:CONFiguration:LIST:STORe
[:SENSe[1]]:CONFiguration:LIST:SIZE?
This command returns the size (number of configuration indexes) of a measure configuration list.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:SIZE? "<name>"
<name>
A string that represents the name of a measure configuration list
Details
This command returns the size (number of configuration indexes) of a measure configuration list. The
size of the list is equal to the number of configuration indexes in a configuration list.
Example
:SENSe:CONF:LIST:SIZE? "MyMeasList"
Returns the number of configuration indexes
in a measure configuration list named
MyMeasList
.
Example output:
3
Also see
(on page 3-37)
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 6-116)