Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-120
DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:CONFiguration:LIST:STORe
This command stores the active measure settings into the named configuration list.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Saved settings
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:STORe "<name>"
[:SENSe[1]]:CONFiguration:LIST:STORe "<name>", <index>
<name>
A string that represents the name of a measure configuration list
<index>
A number that defines a specific configuration index in the configuration list
Details
Use this command to store the active settings to a configuration index in a configuration list. If you do
not include the
<index>
parameter, the configuration index is appended to the end of the list.
Example
:SENSe:CONF:LIST:STOR "MyConfigList"
Stores the active settings of the instrument to the
end of the configuration list named
MyConfigList
.
:SENSe:CONF:LIST:STOR "MyConfigList", 5
Stores the active settings of the instrument to the
configuration list named
MyConfigList
in
configuration index 5.
Also see
(on page 3-37)
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 6-116)