Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 3: Functions and features
DMM7510-901-01 Rev. B / May 2015
3-79
There is a 2 µs delay after the block makes the last measurement in the count.
The measure block must be used with a measure function. The trigger model will not run if a digitize
function is selected.
When you select the measure block, the following options are available.
Setting
Description
Reading Buffer
Optional. The name of the buffer where the measurement is
stored.
Count
Optional. Specifies the number of readings to make before moving
to the next block in the trigger model.
Digitize block
When trigger model execution reaches the block:
1. The instrument begins digitizing measurements.
2. The trigger model execution waits for the measurement to complete.
3. The instrument places the measurements into the specified reading buffer.
If you are defining a user-defined reading buffer, you must create it before you define this block.
When you set the count to a finite value, trigger model execution remains at the block until all digitize
measurements are complete (at least 2 µs). If you set the count to infinite, the trigger model executes
subsequent blocks and digitizing continues in the background until the trigger model execution
reaches another digitize block or until the trigger model ends. The digitize block completes and
moves on to the next block at the end of the aperture time, not the sample rate time.
For example, if there are two readings at a sample rate of 20,000 samples per second (50
µ
s apart)
with an aperture of 1
µ
s, with a delay of 100
µ
s, the delay starts at 51
µ
s. The first reading occurs at 0
µ
s and the second starts at 50
µ
s, but it is completed at 51
µ
s because the aperture is only 1
µ
s. If
the aperture is set to Auto, the first reading is at 0
µ
s, the second starts at 50
µ
s, and the delay starts
at 100
µ
s.
The digitize block must be used with a digitize function. The trigger model will not run if a measure
function is selected.
A trigger model that contains a digitize block may appear to hang in the wait block because it is
making many measurements in one block.
There is a 2
µ
s delay after the digitize block makes the last measurement in the count.
When you select the digitize block, the following options are available.
Setting
Description
Reading Buffer
The name of the buffer where the measurement is stored
Count
Specifies the number of readings to make before moving to the
next block in the trigger model