Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-31
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This command sets the high limit for the limit
Y
test for the selected measurement function. When limit
Y
testing is enabled, the instrument generates a fail indication when the measurement value is more
than this value.
Default is 0.8 for limit 1 when the diode function is selected; 10 when the continuity function is
selected. The default for limit 2 for the diode and continuity functions is 1.
Example
:CALC2:VOLT:LIM1:CLE:AUTO OFF
:CALC2:VOLT:LIM1:AUD FAIL
:CALC2:VOLT:LIM1:LOW 0.25
:CALC2:VOLT:LIM1:UPP 2.5
:CALC2:VOLT:LIMIT1:STAT ON
:READ?
:CALC2:VOLT:LIMIT1:FAIL?
:CALC2:VOLT:LIM1:CLE
Set limit autoclear off.
Enable the beeper for limit 1 when a voltage
measurement exceeds the limit.
Set lower limit 1 for voltage to 0.25 V.
Set upper limit 1 for voltage to 2.5 V.
Enable limit 1 testing for voltage.
Make a reading; the limit is checked and results
display on the front panel
Return the test results; example output if the test
fails on the low limit:
LOW
Clear the test results.
Also see
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA]
:CALCulate2:<function>:LIMit<Y>:STATe
:CALCulate[1]:<function>:MATH:FORMat
This command specifies which math operation is performed on measurements when math operations are
enabled.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
PERC
Usage
:CALCulate[1]:<function>:MATH:FORMat <operation>
:CALCulate[1]:<function>:MATH:FORMat?
<function>
The function to which the setting applies; see
Functions
<operation>
The name of the math operation:
•
y = mx+b:
MXB
•
Percent:
PERCent
•
Reciprocal:
RECiprocal