Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 2: General operation
DMM7510-901-01 Rev. B / May 2015
2-153
To recall a saved setup using TSP commands:
Send the command:
setupName
()
Where
setupName
is the name of the script that contains the setup that was saved with
createconfigscript()
.
Define the setup used when power is turned on
You can select a configuration to be used when power is turned on.
From the front panel:
1. Set the instrument to the settings that you want it to have each time the power is turned on.
2. Press the
MENU
key to open the main menu. Under Scripts, select
Create Setup
. The CREATE
SETUP window is displayed.
3. Select
Create
. A keyboard is displayed.
4. Enter the name of the new script, and then select
ENTER
on the keyboard to save it.
5. The instrument saves all present system settings to the script and displays a confirmation
message. Click
OK
.
6. Press the
EXIT
key to return to the main menu.
7. Under Scripts, select
Run
. The RUN SCRIPTS window opens.
8. Select the script you just created.
9. Select
Copy to Power Up
.
10. Click
OK
on the confirmation message.
Using a SCPI command:
Send the command:
:SYSTem:POSetup <name>
Where
<name>
is:
•
RST
: Use the *RST defaults.
•
SAV0
: Use the setup stored at memory location 0
•
SAV1
: Use the setup stored at memory location 1
•
SAV2
: Use the setup stored at memory location 2
•
SAV3
: Use the setup stored at memory location 3
•
SAV4
: Use the setup stored at memory location 4
Using a TSP command:
Save the script that you want to use as the power-on default to be
autoexec
. For example, to save
the commands that are presently in the instrument to be the power-on defaults, send the command:
createconfigscript("autoexec")
If an
autoexec
script already exists, you must delete it by sending the
script.delete("autoexec")
command. Performing a system reset does not delete the
autoexec
script.