Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-71
dmm.digitize.analogtrigger.edge.level
This attribute defines the signal level that generates the analog trigger event for the edge trigger mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
0
Usage
value
= dmm.digitize.analogtrigger.edge.level
dmm.digitize.analogtrigger.edge.level =
value
value
The signal level that generates the analog trigger event
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This command is only available when the analog trigger mode is set to edge.
The edge level can be set to any value in the active measurement range. See the Model DMM7510
specifications for more information on the resolution and accuracy of the analog trigger.
Example
dmm.digitize.func = dmm.FUNC_DIGITIZE_VOLTAGE
dmm.digitize.analogtrigger.mode = dmm.MODE_EDGE
dmm.digitize.analogtrigger.edge.level = 5
dmm.digitize.analogtrigger.edge.slope = dmm.SLOPE_FALLING
Set the function to digitize voltage.
Set the analog trigger mode to edge.
Set the level to sense 5 V.
Set the level to be detected on a falling edge.
Also see
(on page 3-64)
dmm.digitize.analogtrigger.mode
(on page 8-74)
dmm.digitize.analogtrigger.edge.slope
dmm.measure.analogtrigger.edge.level
(on page 8-118)