Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 2: General operation
DMM7510-901-01 Rev. B / May 2015
2-125
Another example is a 1 k
Ω
/1 k
Ω
/1 k
Ω
resistor network. If 3 V is applied across the total three 1 k
Ω
resistors and V
sense
is across the first 1 k
Ω
resistor, set V
input
to the 10 V range and V
sense
to the 1 V
range. The ratio measurement is approximately 3.00000. If V
sense
is set to the 1 V range, the ratio
displays overflow, with SENSE HI and SENSE LO terminals exceeding the 125 % maximum
reference to the LO terminals. The SENSE HI to LO is 3 V and SENSE LO to LO is 2 V, respectively.
Figure 89: DCV ratio 1 kohm/1 kohm/1 kohm resistor network
Do not apply more than 1000 V
peak
to the INPUT terminals or more than 350 V
peak
to the SENSE
terminals. Failure to heed this caution may result in instrument damage.
DC voltage ratio measure connections
Figure 90: Front panel connections: DC voltage ratio measurement