Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 2: General operation
DMM7510-901-01 Rev. B / May 2015
2-103
Figure 64: Rear-panel connections: AC current measurement (3 A or less)
Figure 65: Rear-panel connections: AC current measurement (3 A or more)
Measure AC current using the front panel
To make an AC current measurement using the front panel:
1. Make the connections as shown in the previous figures.
2. Press the
FUNCTION
key.
3. Select
AC Current
.
4. Press the
TRIGGER
key for 2 seconds and verify that the instrument is set to Continuous
Measurement.
The measurements start displaying on the front panel.
When the TERMINALS switch is set to REAR and autorange is enabled, autoranging is limited to
ranges up to 3 A ranges. The 10 A range is not included in the autorange algorithm.