Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-100
DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:RTD:FOUR
This command contains the type of 4-wire RTD that is being used.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
PT100
Usage
[:SENSe[1]]:<function>:RTD:FOUR <type>
[:SENSe[1]]:<function>:RTD:FOUR?
<function>
The function to which the setting applies; see
Functions
<type>
The type of four-wire RTD:
•
PT100:
PT100
•
PT385:
PT385
•
PT3916:
PT3916
•
D100:
D100
•
F100:
F100
•
User-specified type:
USER
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
The transducer type must be set to temperature and the transducer must be set to 4-wire RTD before
you can set the RTD type.
Example
:FUNC "TEMP"
:TEMP:TRANsducer FRTD
:TEMP:RTD:FOUR PT3916
Set the measure function to temperature.
Set the transducer type to 4-wire RTD.
Set the RTD type to PT3916.
Also see
[:SENSe[1]]:<function>:TRANsducer
(on page 6-112)