Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-156
DMM7510-901-01 Rev. B / May 2015
The options for
<bufferElements>
are described in the following table.
Option
Description
DATE
The date when the data point was measured; not available for reading buffers that
are set to the style compact
EXTRa
Returns an additional value (such as the sense voltage from a DC voltage ratio
measurement); the reading buffer style must be set to full to use this option
FORMatted
The measured value as it appears on the front panel
FRACtional
The fractional seconds when the data point was measured
READing
The measurement reading
RELative
The relative time when the data point was measured
SEConds
The seconds in UTC (Coordinated Universal Time) format when the data point was
measured
STATus
The status information associated with the measurement; see the "Buffer status bits
for sense measurements" table below
TIME
The time when the data point was measured
TSTamp
The timestamp when the data point was measured
UNIT
The unit of measure of the measurement
The
STATus
buffer element returns status values for the readings in the buffer. The status values are
floating-point numbers that encode the status value. Refer to the following table for values.
Buffer status bits for sense measurements
Bit (hex) Name
Decimal Description
0x0001 STAT_QUESTIONABLE
1
Measure status questionable
0x0006 STAT_ORIGIN
6
A/D converter from which reading originated; for the
Model DMM7510, this will always be 0 (Main) or 2
(digitizer)
0x0008 STAT_TERMINAL
8
Measure terminal, front is 1, rear is 0
0x0010 STAT_LIMIT2_LOW
16
Measure status limit 2 low
0x0020 STAT_LIMIT2_HIGH
32
Measure status limit 2 high
0x0040 STAT_LIMIT1_LOW
64
Measure status limit 1 low
0x0080 STAT_LIMIT1_HIGH
128
Measure status limit 1 high
0x0100 STAT_START_GROUP
256
First reading in a group