Appendix 0: Status model
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
4
DMM7510-901-01 Rev. B / May 2015
Bit When set, indicates the following has occurred:
0
Operation complete: All pending selected instrument operations are complete and the instrument is
ready to accept new commands. The bit is set in response to an
(on page 6) command or TSP
(on page 8-230) function.
1
Not used.
2
Query error: Attempt to read data from an empty Output Queue.
3
Not used.
4
Not used.
5
Not used.
6
Not used.
7
Power-on: The instrument has been turned off and turned back on since the last time this register was
read.
You can use the following commands to read and set bits contained in the Standard Event Register.
Description
SCPI command
TSP command
Read the Standard Event Status
Register
(on page 4)
(on page 8-252)
Set or read the OR bits in the
Standard Event Status Enable
Register
(on page 2)
ESE?