Section 2: General operation
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
2-138
DMM7510-901-01 Rev. B / May 2015
Voltage autodelay and autorange times
The following table provides times for autodelay and autorange for the Model DMM7510 DMM
voltage functions.
Function
Detector
bandwidth
Range and delays
DC volts
Not applicable
Range
100 mV 1 V
10 V
100 V
1000 V
Autodelay
1 ms
1 ms
1 ms
5 ms
5 ms
Autorange
10 ms
10 ms
10 ms
50 ms
50 ms
AC volts
Not applicable
Range
100 mV 1 V
10 V
100 V
700 V
3 or 30 Hz
Autodelay
265 ms
265 ms
265 ms
265 ms
1 s
Autorange
2.65 s
2.65 s
7.5 s
7.5 s
10 s
300 Hz
Autodelay
50 ms
50 ms
50 ms
50 ms
250 ms
Autorange
3.5 ms
3.5 s
850 ms
850 ms
2.5 s
Current autodelay and autorange times
The following tables provide times for autodelay and autorange for the Model DMM7510 DMM current
functions.
Function
Range and delays
DC current
Range
10
µ
A
100
µ
A
1 mA
10 mA
100 mA 1 A
3 A
10 A
Autodelay
13 ms
2 ms
2 ms
2 ms
2 ms
2 ms
2 ms
2 ms
Autorange
130 ms
20 ms
20 ms
20 ms
20 ms
20 ms
20 ms
20 ms
Function
Detector
bandwidth
Range and delays
AC current
Not applicable Range
1 mA
10 mA
100 mA
1 A
3 A
10 A
3 or 30 Hz
Autodelay
265 ms
265 ms
265 ms
265 ms
300 ms
300 ms
Autorange
2.65 s
2.65 s
2.65 s
2.65 s
3.0 s
3.0 s
300 Hz
Autodelay
50 ms
50 ms
50 ms
50 ms
75 ms
75 ms
Autorange
500 ms
500 ms
500 ms
500 ms
750 ms
750 ms