Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-4
DMM7510-901-01 Rev. B / May 2015
:MEASure?
This command makes measurements, places them in a reading buffer, and returns the last reading.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:MEASure?
:MEASure:<function>?
:MEASure:<function>? "<bufferName>"
:MEASure:<function>? "<bufferName>", <bufferElements>
:MEASure? "<bufferName>"
:MEASure? "<bufferName>", <bufferElements>
<function>
The function to which the setting applies; see
Functions
<bufferName>
The name of the buffer where the reading is stored; if nothing is specified,
defbuffer1
is used
<bufferElements>
See
Details
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This command makes a measurement using the specified function and stores the reading in a
reading buffer.
If you do not define the function parameter, the instrument uses the presently selected measure
function. If a digitize function is presently selected, an error is generated.
This query makes the number of readings specified by
[:SENSe[1]]:COUNt
. When you use a
reading buffer with a command or action that makes multiple readings, all readings are available in
the reading buffer. However, only the last reading is returned as a reading with the command.
If you define a specific reading buffer, the reading buffer must exist before you make the
measurement.
To get multiple readings, use the
:TRACe:DATA?
command.