Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-263
Example
digio.line[4].mode = digio.MODE_TRIGGER_IN
triggered = trigger.digin[4].wait(3)
print(triggered)
Waits up to 3 s for a trigger to be detected on
trigger line 4, then outputs the results.
Output if no trigger is detected:
false
Output if a trigger is detected:
true
Also see
(on page 8-52)
Digital I/O port configuration
(on page 3-49)
(on page 8-260)
trigger.digout[N].assert()
This function asserts a trigger pulse on one of the digital I/O lines.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.digout[
N
].assert()
N
Digital I/O trigger line (1 to 6)
Details
Initiates a trigger event and does not wait for completion. The pulse width that is set determines how
long the instrument asserts the trigger.
Example
digio.line[2].mode = digio.MODE_TRIGGER_OUT
trigger.digout[2].pulsewidth = 20e-6
trigger.digout[2].assert()
Asserts a trigger on digital I/O line 2 with a pulse
width of 20
µ
s.
Also see
(on page 8-52)
Digital I/O port configuration
(on page 3-49)
(on page 8-264)
(on page 8-264)