dmm.measure.simreftemperature
This attribute sets the simulated reference temperature of the thermocouple reference junction.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
Celsius: 23
Kelvin: 296.15
Fahrenheit: 73.4
Usage
value
= dmm.measure.simreftemperature
dmm.measure.simreftemperature =
value
value
The simulated reference temperature:
•
Celsius: 0 to 65
•
Kelvin: 273 to 338
•
Fahrenheit: 32 to 149
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This attribute applies to the temperature function when the transducer type is set to thermocouple and
the reference junction is set to simulated. It allows you to set the simulated temperature value.
Example
dmm.measure.func = dmm.FUNC_TEMPERATURE
dmm.measure.transducer = dmm.TRANS_THERMOCOUPLE
dmm.measure.unit = dmm.measure.UNIT_CELSIUS
dmm.measure.simreftemperature = 30
Sets 30 degrees Celsius as the simulated
reference temperature for thermocouples.
Also see