![Keithley DMM7510 Скачать руководство пользователя страница 507](http://html2.mh-extra.com/html/keithley/dmm7510/dmm7510_reference-manual_4027507507.webp)
Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-180
DMM7510-901-01 Rev. B / May 2015
:TRIGger:BLENder<n>:STIMulus<m>
This command specifies the events that trigger the blender.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Trigger blender clear
Save settings
NONE
Usage
:TRIGger:BLENder<n>:STIMulus<m> <event>
:TRIGger:BLENder<n>:STIMulus<m>?
<n>
The blender number (1 or 2)
<m>
The stimulus input number (1 to 4)
<event>
See
Details
Details
There are four stimulus inputs that can each select a different event.
Use zero to disable the blender input.
The
<event>
parameter may be any of the trigger events shown in the following table.
Trigger events
Event description
Event constant
No trigger event
NONE
Front-panel TRIGGER key press
DISPlay
Notify trigger block
<n>
(1 to 8); the trigger model generates a
trigger event when it executes the notify block
NOTify<n>
A command interface trigger (bus trigger):
•
Any remote interface: *
TRG
•
GPIB only: GET bus command
•
VXI-11: VXI-11 command
device_trigger
COMMand
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
<n>
(1 to 6)
DIGio<n>
Line edge detected on TSP-Link synchronization line
<n>
(1 to 3)
TSPLink<n>
Appropriate LXI trigger packet is received on LAN trigger object
<n>
(1 to 8)
LAN<n>
Trigger event blender
<n>
(1 or 2), which combines trigger events
BLENder<n>
Trigger timer
<n>
(1 to 4) expired
TIMer<n>
Analog trigger
ATRigger
External in trigger
EXTernal