Section 2: General operation
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
2-24
DMM7510-901-01 Rev. B / May 2015
DC voltage measure settings
The following options are available on the Measure Settings menu when the function is set to DC
voltage.
Setting
Description
Range
Determines the full-scale input for the measurement; also affects
the accuracy of the measurements and the maximum signal that
can be measured. Refer to
(on page 3-3).
Auto Delay
Applies a wait period at the start of measurement to allow cables
and circuitry to settle for best accuracy. Refer to
Display Digits
Sets the number of digits that are displayed for front-panel
readings. It does not affect accuracy or speed. Refer to
(on page 2-55).
Input Impedance
Sets impedance to Auto or 10 M
Ω
. Refer to
(on page 2-98).
Unit
Allows voltage to be measured in volts or decibels. Refer to
(on page 2-98).
Integration Rate
Controls the amount of time the input signal is measured
(aperture), which affects the noise and reading rate; see
aperture or NPLCs to adjust speed and accuracy
(on page 4-1).
Auto Zero
Determines if internal reference points are used to maintain
stability and accuracy. See
Automatic reference measurements
Count
Sets the number of aperture readings that are processed when a
measurement is requested.
Line Sync
Enables or disables line synchronization. When it is enabled, it
helps increase common-mode and normal-mode noise rejection.
Refer to
Decibel Reference
Sets decibel reference point; this setting is only available when Unit
is set to Decibel. Refer to
Show voltage readings in decibels