Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-33
bufferVar.readings
This attribute contains the readings stored in a specified reading buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Restore configuration
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
reading
=
bufferVar
.readings[
N
]
reading
The value of the reading in the specified reading buffer
bufferVar
The name of the reading buffer, which may be a default buffer (
defbuffer1
or
defbuffer2
) or a user-defined buffer
N
The reading number
N
; can be any value from 1 to the number of readings in the
buffer; use the
bufferVar.
n
command to determine the number of readings in the
buffer
Example
reset()
testData = buffer.make(50)
trigger.model.load("SimpleLoop", 3, 0, testData)
trigger.model.initiate()
waitcomplete()
printbuffer(1, 3, testData.readings)
Create a reading buffer named
testData
, configure the instrument to make three measurements, and store
the readings in the buffer.
Print the three readings in
testData
.
Output:
-9.6420389034124e-12, -4.5509945811872e-10, -9.1078204006445e-12
Also see
(on page 8-32)
(on page 8-16)
(on page 8-18)
(on page 8-24)
(on page 8-231)
(on page 8-232)
(on page 3-13)