Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-29
:CALCulate2:<function>:LIMit<Y>:STATe
This command enables or disables a limit test on the measurement from the selected measure function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0 (OFF)
Usage
:CALCulate2:<function>:LIMit<Y>:STATe <state>
:CALCulate2:<function>:LIMit<Y>:STATe?
<function>
The function to which the setting applies; see
Functions
<Y>
Limit number:
1
or
2
<state>
Disable the limit test:
OFF
or
0
Enable the limit test:
ON
or
1
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This command enables or disables a limit test for the selected measurement function. When this
attribute is enabled, the limit
Y
testing occurs on each measurement made by the instrument. Limit
Y
testing compares the measurements to the high and low limit values. If a measurement falls outside
these limits, the test fails.