Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-161
dmm.measure.limit[Y].enable
This attribute enables or disables a limit test on the measurement from the selected measure function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
dmm.OFF
Usage
state
= dmm.measure.limit[
Y
].enable
dmm.measure.limit[
Y
].enable =
state
state
Limit
Y
testing:
•
Disable:
dmm.OFF
•
Enable:
dmm.ON
Y
Limit number:
1
or
2
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This command enables or disables a limit test for the selected measurement function. When this
attribute is enabled, the limit
Y
testing occurs on each measurement made by the instrument. Limit
Y
testing compares the measurements to the high and low limit values. If a measurement falls outside
these limits, the test fails.
Example
See
dmm.measure.limit[Y].low.value
(on page 8-164) for an example of how to use this command.
Also see
(on page 8-95)
dmm.measure.limit[Y].autoclear
(on page 8-159)
(on page 8-160)
dmm.measure.limit[Y].high.value
(on page 8-163)
dmm.measure.limit[Y].low.value
(on page 8-164)