LOGIQ™ A3 S
ERVICE
M
ANUAL
Chapter 10
10-3
Table 10-2 Customer Care Schedule
Service at Indicated Time
Daily
Weekly
Monthly
Annually
Notes
Clean Probes
•*
* or before each use
Clean Probe Holders
•
Clean Air Filter
•
more frequently depending on
your environment
Inspect AC Mains Cable
•
Mobile Unit Check Weekly
Inspect Cables and Connectors
•
Clean Console
•
Clean Monitor
•
Inspect Wheels, Casters, brakes and
Swivel Locks
•
Mobile Unit Check Daily
Console Leakage Current Checks
•
also after corrective
maintenance
Peripheral Leakage Current Checks
•
also after corrective
maintenance
Surface Probe Leakage Current Checks
•
also after corrective
maintenance
Endocavity Probe Leakage
Current Checks
Twice
Annually
Surgical Probe Leakage
Current Checks
As Prescribed
in probe manual
Measurement Accuracy Checks
•
also after corrective
maintenance