Section 8-5 - I/O Interfaces
8-5-2
CA9-Transformer to Front Panel and power supply cable assy (FRU P/N #
5192656)
8-5-2-1
Tools
•
Common phillips screwdrivers
8-5-2-2
Needed Manpower
•
1person, 30 m travel
8-5-2-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-5-2-4
Removal Procedure
1. Remove LH and RH plastic covers.
2. Remove two screws of transformer.
3. Disconnect J4 from TRX.
4. Remove LH and RH EMI Covers.
5. Remove Card Cage EMI Cover.
6. Slide out the HVLV and disconnect the power input cable for HVLV.
7. Remove SSR by removing 2 fasteners and disconnect cable from fan assy.
8. Remove the cable from all the cable clamps
9. Remove the Top Cover
10. Cut the required cable ties
11. Remove Keyboard Rear Cover
12. Remove front panel
13. Remove the grounding cables.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 197 of 252