Chapter 8
8-29
8-4-4
Front Cover (FRU No. 5271834)
This is a description on how to remove and replace the Front Cover.
8-4-4-1
Tools
•
Common Phillips screwdrivers
8-4-4-2
Needed Manpower
•
1 persons, 15 m travel
8-4-4-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-4-4-4
Removal Procedure
for details on how to remove Right cover.
2.) Remove Left cover.Refer
for details on how to remove Left cover.
3.) Loosen the four screws (a, b, c, d)
4.) Remove the front cover..
8-4-4-5
Functional Checkout Procedure
Figure 8-23 Front Cover
See Section
Functional and/or
Leakage Current
Test
Debrief Script
Service Manual Direction 5213326-100, Rev 3, Section 8-4-4. Equipment passes all
required tests and is ready for use.
Basic Controls
B- Mode Checks
M-Mode Checks
a
b
c
d
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.