Section 4-4 - Functional Checks
4-4-6
Probe/Connectors Usage
4-4-6-1
Connecting a probe
1.) Place the probe's carrying case on a stable surface and open the case.
2.) Carefully remove the probe and unwrap the probe cable.
3.) DO NOT allow the probe head to hang free. Impact to the probe head could result in irreparable
damage.
4.) Turn the connector locking handle counterclockwise.
5.) Align the connector with the probe port and carefully push into place.
6.) Turn the connector locking handle clockwise to secure the probe connector.
7.) Carefully position the probe cable in the probe cord holder spot so it is free to move, but not
resting on the floor.
4-4-6-2
Activating the probe
Press
“Probe”
Key available on the control panel. The next available probe will be actiavetd and
presets will be loaded automatically.
When deactivating the probe, the probe is automatically placed in standby mode.
1.) Press the
Freeze
key.
2.) Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual for
complete probe cleaning instructions.)
3.) Carefully slide the probe around the right side of the keyboard, toward the probe holder.
Ensure that the probe is placed gently in the probe holder.
4-4-6-3
Disconnecting the probe
Probes can be disconnected at any time. However, the probe should not be selected as the active
probe.
1.) Move the probe locking handle counterclockwise. Pull the probe and connector straight out of
the probe port.
2.) Carefully slide the probe and connector away from the probe port and around the right side of
the keyboard. Ensure the cable is free.
3.) Be sure that the probe head is clean before placing the probe in its storage box.