Section 8-4 - Mechanicals
Section 8-4 Mechanicals
8-4-1
RH Side Cover (FRU No. 5265162)
This is a description on how to remove and replace the Right Cover.
8-4-1-1
Tools
•
Common Phillips screwdrivers
8-4-1-2
Needed Manpower
•
1person, 15 m travel
8-4-1-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-4-1-4
Removal Procedure
1.) Unscrew the two screws (1-2).
2.) Remove the Right cover in the direction as shown in the figure below.
Figure 8-20 Right Cover
8-4-1-5
Functional Checkout Procedure
See Section
Functional and/or
Leakage Current
Test
Debrief Script
Service Manual Direction 5213326-100, Rev 3, Section 8-4-1. Equipment passes all
required tests and is ready for use.