GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
8-42
Section 8-4 - Mechanicals
8-4-14
Fan Assy (FRU No. 5215345)
This is a description on how to remove and replace the Card Cage Fan Assy.
8-4-14-1
Tools
•
Common Phillips screwdrivers
8-4-14-2
Needed Manpower
•
1 persons, 15 m travel
8-4-14-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-4-14-4
Removal Procedure
1.) Remove the Right Cover. Refer to section
2.) Remove the Right EMI Cover. Refer to section
3.) Disconnect the Fan Assembly cable from the backplane.
4.) Remove SSR cables on FAN Assembly.
Figure 8-35 Removing Fan cable
Open here
Open here
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 191 of 252