GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
8-4
Section 8-2 - LCD
8-2-2
LCD Arm Cover Set (FRU P/N # 5270335)
8-2-2-1
Tools
•
Common phillips screwdrivers
8-2-2-2
Needed Manpower
•
1 person, 15 m travel
8-2-2-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-2-2-4
Removal procedure
1. Remove Keyboard Assembly.
2. Remove Keyboard rearcover by unfastening the 2 screws holding it.
3. Remove-LCD-Arm-Rear-Cover assy.
4. Remove Monitor Neck Cover by removing the 4 screws holding it and disconnect the USB cable from
inside
Figure 8-4 LCD Arm Rear Cover and Monitor Neck Cover
LCD Arm Rear
Cover
Monitor Neck Cover
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.