Section 2-2 - General Console Requirements
Section 2-2General Console Requirements
2-2-1
Console Environmental Requirements
2-2-1-1
Cooling
The cooling requirement for the LOGIQ A3 is 1366 BTU/hr. This figure does not include cooling needed
for lights, people, or other equipment in the room. Each person in the room places an additional 300
BTU/hr. demand on the cooling system.
2-2-1-2
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
Table 2-2
Environmental Requirements for LOGIQ A3 Scanners
Operation
Storage
Transport
Temperature
10 - 40 degree C
50 - 104 degree F
-10 - 60 degree C
14 - 140 degree F
-40 - 60 degree C
-40 - 140 degree F
Humidity
30 - 85%
non-condensing
30 - 85%
non-condensing
30 - 85%
non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
Table 2-3
Environmental Requirements for an Ultrasound Room
Item
Values
Power Source
Refer to Table 2-4 on page 2-3.
Radiation Shielding
NONE REQUIRED for ULTRASOUND ENERGY
Temperature
20-26 DEG. C (68-79 DEG F) for PATIENT COMFORT
Humidity
50% to 70% for PATIENT COMFORT
Heat Dissipation
1366 BTU/Hr.
Floor Landing
Approximately 680 - 800 kg/m
2
without Accessories
Floor Condition
Gradient: WITHIN 5 degrees
Weight
70 kg without Accessories