Chapter 7
7-17
7-5-9
Peripheral Power on Top Panel absent
Peripheral Power on Top
Panel absent
Connect mains cable and
Switch ON the system.
Ensure Circuit breaker is
switched ON
Check if the fuses for
Peripherals on the
Transformer module are not
blown off.
Replace Transformer Module
if it is blown off
Check the Front Panel cable
from Transformer Module to
the front panel.
Check and ensure the SSR
relay is not defective, Replace
if found defective