GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
8-2
Section 8-2 - LCD
Section 8-2 LCD
8-2-1
LCD ASSEMBLY(FRU P/N # 5271538)
This is a description on how to remove and replace the LCD Assembly.
8-2-1-1
Tools
•
Common phillips screwdrivers
8-2-1-2
Needed Manpower
•
1 person, 15 m travel
8-2-1-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-2-1-4
Removal procedure
1.) Flip down the LCD Monitor.
2.) Unscrew the two screws (5, 6) and remove LCD arm cover, Refer
Figure 8-1 LCD Rear Cover
3.) Unscrew the four screws (1,2,3,4) to remove LCD Rear Cover, Refer
4.) Unfasten the cable clamp for Power and Signal cables and disconnect the cable connection to
the LCD
Figure 8-2 Remove Cables
1
2
3
4
5
6
LCD Rear Cover
LCD Arm Rear Cover
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 151 of 252