Chapter 7
7-21
7-5-13
Trackball malfunction
Trackball malfunction
Is A/N and
Ultrasound
Keyboard test
ok ?
Y
Check and confirm the
Trackball test.
Replace Trackball FRU part if
found defective
Enter 'Service Diagnostics'
from 'Utility' screen.
Enter login id and password.
Run Keyboard Tests
Check all Keyboard Tests
including Trackball tests
Checkand replace the entire
Keyboard Asm