GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
8-60
Section 8-6 - PCB
8-6-2
Connector Board Assy (FRU No. 5193606)
This is a description on how to remove and replace the Conn. Board Assy.
8-6-2-1
Tools
•
Common Phillips screwdrivers
8-6-2-2
Needed Manpower
•
1 person, 15 m travel
8-6-2-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-6-2-4
Removal Procedure
1.) Remove Right Cover. Refer
for details on how to remove Right Cover.
2.) Remove Left Cover. Refer
for details on how to remove Left cover.
3.) Remove Front Cover. Refer
for details on how to remove front cover.
4.) Remove Front EMI Cover. Refer
for details on how to remove front EMI cover.
5.) Unscrew the four screws (1-4) from the connector board
6.) Disconnect the connector board assembly.
CAUTION
An electronic discharge may damage a component. Turn OFF power and wear the wrist strap
before you remove circuit boards. Do not unplug the power cord to keep ground continuity.
Do not bend or flex the boards when mounting/dismounting each boards. Surface mount IC
boards are very susceptible to damage from flex/torque.
Figure 8-50 Removing Connector Board Assembly
1
2
3
4
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 209 of 252