Section 5-2 - Block Diagrams and Theory
5-2-4
Power Supply(LV/HVPS)
An integrated power supply (Low voltage & High voltage) is required to supply all required voltages to
Backend and Front End hardware of LOGIQ Uno. Following are the voltage requirements for the Uno
hardware:
Table 6:
PCB Name
Voltage Source
Max. Current
Tolerance/
Range
Connector/ Routing
CONN Board
+5V
1.25Amp
±5%
Directly from integrated
power supply through
BPB
-5V
10mAmp
±5%
--
12VA
20mAmp
±5%
--
+SHV
TBD
+80V
--
-SHV
TBD
-80V
--
TxRx Board
+5V
1.3Amp
±5%
Directly from integrated
power supply through
BPB
+3.3V
2Amp
±5%
--
12VA
250mAmp
±5%
--
+THV
TBD
+2.5V to +60V
--
-THV
TBD
-2.5V to –60V
--
MST Board
+3.3V
2Amp
±5%
Directly from integrated
power supply through
BPB
+5V
450mAmp
±5%
--
-5V
140mAmp
±5%
--
12VA
90mAmp
±5%
--
FAN
12VA
400mAmp
±5%
Directly from integrated
power supply through
BPB
USB-HDD
+5V
1Amp
±5%
--
LCD
12VA
TBD
±5%
--