Section 7-5 - Troubleshooting Guide
7-5-2
No Ultrasound Image
No Ultrasound Image
Switch ON the system.
Confirm the system boots up to the
echoloader Image screen
Check and Confirm the probe is
recognized. Check and confirm the
probe is not defective. Also, check
and confirm if Image is obtained
on second probe port.
Enter 'Service Diagnostics' from
'Utility' screen.
Enter login id and password.
Run 'All' board Tests
Is Defective
identified?
Replace the identified
defective board
( CON/ TxRx)
Y
Check and confirm about presence
of HV 'symptom' on Image near
field.
Also check error and event logs for
the HV voltages
N
Is HV symptom
available?
Check and Replace the
HV board
N
Y
Confirm all boards including Digital
Interface board are properly connected.
If issue persists, replace TxRx and MST
boards