GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
Chapter 8
8-15
8-3-4
Optical Trackball Assy (FRU P/N # 5262338)
8-3-4-1
Tools
•
Common phillips screwdrivers
8-3-4-2
Needed Manpower
•
1 person, 10 m travel
8-3-4-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-3-4-4
Removal procedure
1. Remove Keyboard Assembly.
2. Unfasten the 3 screws holding the trackball bracket (circled in cyan color in Figure 8-12)
3. Remove Trackball assembly with Trackball Interface cable.
Figure 8-12 Optical Trackball Assembly
8-3-4-5
Functional Checkout Procedure
See Section
Functional and/or
Leakage Current
Test
Debrief Script
Service Manual Direction 5213326-100, Rev 3, Section 8-3-4. Equipment passes all
required tests and is ready for use.
Basic Controls
B- Mode Checks
M-Mode Checks
Measurement Check
Peripheral Checks
Service Diagnostics
3 screws to be
removed
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 164 of 252