8-4-8
Front Bumper (FRU No. 5168362)
This is a description on how to remove and replace the Front Bumper.
8-4-8-1
Tools
•
Common Phillips screwdrivers
8-4-8-2
Needed Manpower
•
1 persons, 15 m travel
8-4-8-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-4-8-4
Removal Procedure
1. Remove Front Cover. Refer
for details on how to remove Front Cover
2. Remove the two screws (marked 1, 2 in Figure 8-28)
3. Remove Front Bumper
8-4-8-5
Functional Checkout Procedure
Figure 8-28 Removing Front Bumper
See Section
Functional and/or
Leakage Current
Test
Debrief Script
Service Manual Direction 5213326-100, Rev 3, Section 8-4-8. Equipment passes all
required tests and is ready for use.