GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
8-14
Section 8-3 - Keyboard
8-3-3
Alpha Numeric Keyboard with cables (FRU P/N # 5181790)
8-3-3-1
Tools
•
Common phillips screwdrivers
8-3-3-2
Needed Manpower
•
1 person, 30 m travel
8-3-3-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-3-3-4
Removal procedure
1. Remove Keyboard Assembly.
2. Remove A/N Keyboard cable
3. Remove the keyboard top hinge by unfastening the 2 screws marked A & B in Figure 8-11
3. Remove the Alphanumeric Keyboard by unfastening the 10 screws (marked in cyan color in Figure
4-11) holding it
Figure 8-11 A/N Keyboard PCB
8-3-3-5
Functional Checkout Procedure
See Section
Functional and/or
Leakage Current
Test
Debrief Script
Service Manual Direction 5213326-100, Rev 3, Section 8-3-3. Equipment passes all
required tests and is ready for use.
Basic Controls
B- Mode Checks
M-Mode Checks
Measurement Check
Peripheral Checks
Service Diagnostics
10 Screws
A/N Keyboard PCB
B
A
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 163 of 252