Chapter 3
3-13
3-4-1-2
Weight with Monitor and without Peripherals
The Weight of LOGIQ A3 with monitor, without peripherals
3-4-1-3
Acoustic Noise Output:
Less than 70dB (A) according to DIN 45635 - 19 - 01 - KL2.
3-4-2
Electrical Specifications
Electrical conduit, junction boxes, outlets, circuit breakers, and switches should be in place before
installing the LOGIQ A3 console
3-4-3
Probe (Transducer) Connection
1.) Connect a transducer to the upper transducer receptacle as follows:
A.) Ensure that the transducer twist lock lever points towards the 9 o'clock position.
B.) Insert the transducer connector on the receptacle guide pin until it touches the receptacle
mating surface.
C.) Twist the transducer twist lock lever to the 2 o'clock position to lock it in place. Twist the
lever to the 9 o'clock position to disconnect the transducer.
NOTE:
It is not necessary to turn OFF power to connect or disconnect a transducer.
Connect the main power cable to a hospital grade power receptacle with the proper rated voltage
checked during pre installation. Never use a three-to-two prong adapter; this defeats the safety ground.
Table 3-5
Weight of LOGIQ A3
Model
Weight [Kg]
Weight [lbs]
LOGIQ A3
70
154
Table 3-6 Electrical Specification for LOGIQ A3
System
Voltage
Current
Frequency
1
120 ~ VAC
4.25A
50-60 Hz
2
230 ~ VAC
1.94A
50-60 HZ