Section 8-3 - Keyboard
8-3-2
TGC PWA FRU Assy (FRU P/N # 5266629)
8-3-2-1
Tools
•
Common phillips screwdrivers
8-3-2-2
Needed Manpower
•
1 person, 10 m travel
8-3-2-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-3-2-4
Removal procedure
1. Remove keyboard Assembly.
2. Remove TGC Caps. ( 6 Nos.)
3. Remove TGC assembly by removing the 4 screws (as marked in figure 4-10) PWA with TGC interface
cable.
Figure 8-10 TGC Assembly