Calibration
redox relative
"Data entry absolute"
The transmitter has a calibrated mV display range.
One
absolute mV value is set with
a single buffer solution (adaptation of the measuring chain offset). A buffer solution
preferably with 225 or 475 mV is used.
Calibration
redox relative
"Calibration absolute"
The transmitter has a calibrated mV display range.
One
absolute mV value is set with
a single buffer solution (adaptation of the measuring chain offset). A buffer solution
preferably with 225 or 475 mV is used.
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
CC
Calibration with
abs. data entry
Note in display
Display of the type of on-site calibration selected in the
calibration settings.
–
CCA1
0000 mV
(–1500 ... +1500 mV)
Entry of offset value
Enter the mV value for the electrode offset (electrode
offset = deviation of the measured value display from
buffer solution mV value)
Confirm by pressing "E". The entered value is effective
immediately.
The maximum offset is 400 mV.
E2
CCA3
Offset too high / too low
Note in display:
Error message if the entered offset leaves the
maximum range.
–
CCA4
Accept
Reject
Recalibrate
End of calibration
Accept
: Pressing "E" accepts the new calibration data.
Reject
: The data is not accepted, a recalibration won’t
be started.
Recalibrate
: The data is rejected and a new calibration
will be started.
E1
CCA5
Electrode in medium?
Note in display:
If the electrode is back in the medium, so that
measurement can take place?
–
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
CC
Calibration with
abs. calibration
Note in display
Display of the type of on-site calibration selected in the
calibration settings.
–
CCB1
Immerse:
Electrode in buffer
Handling instructions
Immerse the electrode in the buffer and then press "E".
–
CCB2
0225 mV
(–1500 ... +1500 mV)
Enter buffer
During calibration, enter the buffer mV value.
E2
CCB3
Time: 10 s
mV 1:
225
Checks the stability of the calibration
Wait until the measurement is stable:
Time does not count down,
mV value no longer flashes,
Display "Measured value stable"
Then press "E".
–
CCB4
Invalid calibration value
Note in display:
Error message if the entered offset is too large.
–