TopClean S CPC 30
8 Trouble-shooting
Hauser
101
E080
F
U
N
C
T
I
O
N
C
H
E
C
K
Range for current
output 1 too small
Increase measuring range span for current output
assignment
no
no
no
E081
Range for current
output 2 too small
no
no
no
E090
CPG 30 service switch
active
Check on CPG if service was actually carried out.
no
no
no
E100
Current simulation
active
Check if functions were consciously selected.
no
no
no
E101
Service function active
no
no
no
E106
Download active
Wait for download to end.
no
no
no
E116
Download error
Repeat download.
no
no
no
E152
PCS Alarm
pH sensor defective or totally soiled; measured water
flow in bypass interrupted;
air cushion in assembly; measuring line interrupted.
no
no
no
E156
Calibration timer run
out
Time for calibration!
no
no
no
E164
Dynamic range of pH
convertor exceeded
Check cable / sensor.
no
no
–
E166
Dynamic range of
reference convertor
exceeded
no
no
–
E168
SCS message IsFET
sensor 1
Leak current > 200 nA. Early warning. Work can be
continued until error E008/E009 occurs.
no
no
–
Error
No.
N
A
M
U
R
Error message
Possible causes / measures
Alarm
contact
Error
current
Automatic
cleaning
start
Fact.
User
Fact.
User
Fact.
User