4 Wiring
TopClean S CPC 30
24
EHauser
4.4
Post-connection check
Perform the following checks after completing electrical installation of the measuring
device:
Instrument status and specifications
Remarks
Is the measuring instrument or the cable damaged externally?
Visual inspection
Electrical connection
Remarks
Does the supply voltage match the specifications on the nameplate?
CPM 153:
100 V ... 230 V AC long-range
24 V AC / DC
CPG 30:
110/230 V AC recodeable
24 V AC / DC
Do the cables used fulfil the required specifications?
Use a genuine E+H cable for
electrode and sensor connection,
see chapter "Accessories".
Are the installed cables strain-relieved?
Is the cable type route completely isolated?
Along the whole cable length, run
the power supply and signal line
cables separately to avoid any
mutual influence. Cable channels
are best.
No loops and cross-overs in the cable run?
Are the power supply and signal cables correctly connected?
Are all the screw terminals properly tightened?
For connection with potential matching (PML):
Is the PML connected to the measuring medium or the buffer
solution?
!
Note!
During calibration, insert the PML
into the buffer solution.
Are all the cable entries installed, tightened and sealed?
Cable run with "water sag"?
"Water sag": cable circuit hanging
down so that water can drip off.
Are all the housing covers installed and tightened?
Check seals for damage.