TopClean S CPC 30
6 Commissioning
Hauser
75
Here, you can activate/deactivate the five limit contactors and define their limits within
which an error message is output.
PARAM
⇒
Controller values
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
Q1
Function
On-value:
Off-value:
off
pH 16.00
pH 16.00
Selection limit contactor 1 / 2 / 3 / 4 / 5
Function:
on/off. The limit contactor is activated/
deactivated.
On-value
: The lowest measured value at which an error
message shall be output.
Off-value: The highest measured value at which an
error message shall be output.
E1
PARAM
⇒
Manual operation
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
R1
TopClean S
ChemoClean
Hold
Contacts
Select manual operation
!
Note!
• Leave the manual operating menu by pressing
"PARAM", "DIAG" or "MEAS".
• The settings are only active in this menu. Nothing is
saved when you leave.
E1
R2
!!!Caution!!
You are now leaving manual
operation.
Note in display (no entry):
–
TopClean S:
RA1
Weekly progr.
Clean trigger
Ext. control
off
off
off
Note in display (no entry):
System status
–
!
Note!
Any
external
programme start is suppressed in the grey menu area below.
RA2
Move assembly
Prog. start
Prog. stop
Selection
You can move the assembly manually or start/stop a
programme.
E1
Move assembly:
RAA1
Ass. service
Ass. measurement
Select position
into which the assembly should move.
E1
RAA2
Weekly progr.
Clean trigger
Ext. control
off
off
off
Note in display (no entry):
System status
–