10 Technical data
Hauser
117
10
Technical data
10.1
Input
Measured variables
pH, redox, temperature
pH (glass / IsFET)
Redox
Temperature
Digital inputs
Mycom S CPM 153:
Measuring range
–2.00 ... +16.00
Measured value resolution
pH 0.01
Zero shift range
pH –2 ... +16
Range of automatic temperature compensation
–50 ... +150 °C
Reference temperature
25 °C (settable with
medium temperature
compensation)
Slope adjustment
5 ... 99 mV / pH
Input impedance under nominal operating conditions > 1 · 10
12
Ω
Input current under nominal operating conditions
< 1.6 · 10
–12
A
Measuring range
–1500 ... +1500 mV
–3000 ... +3000%
Measured value resolution
0.1 mV
Zero shift range
+200 ... –200 mV
Assignment with % display
adjustable,
∆
for 100 % = 150 ... 2000 mV
Electrode offset
±120 mV
Input impedance under nominal operating conditions > 1 · 10
12
Ω
Input current under nominal operating conditions
< 1.6 · 10
–12
A
Temperature sensor
Pt 100 (three-wire circuit)
Pt 1000
NTC 30k
Measuring range (can also be displayed in °F)
–50 ... +150°C
(NTC: –20 ... 100 °C)
Measured value resolution
0.1 K
Temperature offset
± 5K
Input voltage
10 ... 50 V
Internal resistance
R
i
= 5 k
Ω