6 Commissioning
TopClean S CPC 30
62
EHauser
Fig. 27: Diagram of the relationship between switch-on and switch-off points and on and off delay
Switch-on point
Switch-on point
Switch-off point
Switch-off point
Alarm threshold
Measured value
Contact ON
Contact ON
Alarm ON
Alarm OFF
Contact OFF
Contact OFF
Switch-on point > switch-off point:
Switch-on point < switch-off point:
C07-CPM153xx-05-06-00-en-006
on
delay
off
delay
alarm
delay
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
P1
Limit contactor 1
Limit contactor 2
Limit contactor 3
Limit contactor 4
Limit contactor 5
Selection
of the limit contactor which you wish to configure. There
are five limit contactors available.
E1
Limit contactor 1 / 2 / 3 / 4 / 5:
PA1 / PB1 / PC1 / PD1 / PE1
Function
Assignment
On point:
Off point:
off
pH/Redox
pH 16.00
(1500 mV/
100 %/150°C)
pH 16.00
(1500 mV/
100 %/150°C)
Limit contactor configuration:
Function
: Activation of function as limit contactor
Assignment
: Selection of the measured value which
shall valid for the limit value (pH/Redox K1,
temperature)
On point
: Entry of the value at which the limit function is
activated.
Off point:
Entry of the value at which the limit function is
deactivated.
(Settable range: pH –2.00 ... 16.00 /
–1500 mV ... +1500 mV / 0 ... 100% / –50 ... +150 °C)
E2
PA2 / PB2 / PC2 / PD2 / PE2
On delay:
Off delay:
Alarm
threshold:
0000 s
0000 s
pH 16.00
(150°C)
Limit contactor configuration:
On delay
.: Entry of the switch-on delay
(range 0 ... 2000 s)
Off delay
:Entry of the switch-off delay
(range 0 ... 2000 s)
Alarm threshold
: Entry of the value (alarm threshold) at
which the fault signalling contact switches.
E2