6 Commissioning
TopClean S CPC 30
38
EHauser
PARAM
⇒
Set up 1
⇒
Contacts
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
F1
NAMUR:
Relay:1
Relay:2
off
free
free
Contact functions
Depending on the equipment available, you can assign
the function of up to 5 relays here. If you switch on
NAMUR, relays 1 and 2 are assigned and are not
available for other functions (compare page 19).
Selection:
Free / Controller / LC / CCW / CCC
Controller:
Controller control using relay
LC
: Limit contactor function
CCW
: ChemoClean water. Water supply for the
ChemoClean function.
CCC
: ChemoClean Cleaner. Cleaner supply for the
ChemoClean function.
(Together CCC and CCW form the "ChemoClean"
function; You can find information on ChemoClean on
page 71)
The limit value/controller contacts are configured in the
menu "PARAM"
➞
"Set up 2"
➞
"Controller settings".
!
Note!
•
Danger of data loss.
If the controller is already
completely configured for output via relays and you
reduce the number of relays assigned to a controller,
then the
complete
controller configuration (s.
page 50) is reset to the default values.
• If you change the relay assignment for the
controllers, you must use the controller menu
(s. page 50) to reassign a relay to all the functions
selected there.
Example
: Relays 4 and 5 are assigned to the
controller and you change the controller assignment
to relays 5 and 6 (number of relays remains 2)
(no data loss, providing the number of assigned
relays is not reduced!).
• You can only activate NAMUR if the required relays 1
and 2 (compare page 19) are free.
E2
If the NAMUR function is enabled, the Alarm, Relay 1 and Relay 2 contacts are given the following functions:
"Failure" = Fault-signaling contact (terminals 41/42): Failure alarms are active if the measuring system is not working correctly or if process
parameters have reached a critical value.
"Maintenance required " = Relay 1 (Terminals 47/48): Warning messages become active when the measuring system is working correctly but
requires maintenance or a process parameter has reached a value which requires intervention.
"Function check" = Relay 2 (Terminals 57/58): This contact is active during calibration, maintenance, configuration and during the automatic
cleaning/calibration cycle.
F2
NC contact
NO contact
Selection acc. to NAMUR:
Assignment of NAMUR contacts as NC contact or NO
contact.
E1
NC contact = contact open when relay active
NO contact = contact closed when relay active
F3
NC contact
NO contact
Select controller contacts as NC contact or NO
contact.
E1
F4
NC contact
NO contact
Select limit values as NC contact or NO contact.
E1